Skip to content

S3A portable suspended solids analyser for activated sludge.

The S3A is a portable MLSS analyser and suspended solids settlement analyser for routine activated-sludge checks. It measures MLSS from a representative 210 ml sample, then records SVI, settled sludge level, supernatant clarity and floating sludge during a 30-minute cycle. Compare suspended solids analyser roles or review the portable MLSS application.

S3A analyser on a countertop overlooking a modern water-recovery campus

Portable MLSS analysis and settlement testing in one system.

Prepare a representative activated-sludge sample, fill the S3A tube and start the test. The analyser manages the MLSS measurement and settlement cycle, then stores the result at the unit.

01 / MLSS

MLSS from a compact sample

The S3A measures mixed liquor suspended solids from a compact 210 ml sample and provides a result in approximately 10 seconds. The specified MLSS range is 50 to 6,500 mg/L, with stated accuracy of ±10%. Readings above 6,500 mg/L are indicated as INF.

  • A rapid operational MLSS result

  • Suitable for mixed liquor and suspended-solids process checks

  • Specified MLSS accuracy of ±10%

02 / SETTLEMENT TEST

Unattended settlement analysis

After the rapid MLSS result, the S3A continues the 30-minute settlement cycle unattended. It measures settled sludge level, calculates SVI and records supernatant clarity and floating sludge from the same sample.

  • Automated 30-minute settlement cycle

  • Supports routine activated-sludge process checks

  • One sample. Five essential measurements.

Settlement analysis brought together in one instrument

Operators can start the test, continue with other work and return after completion to review automatically calculated results.

210 ml

Compact representative sample

≈10 sec

MLSS measurement

30 min

Unattended settlement test

28

Completed results per location

3

Stored locations

2,000–4,000 mg/L

Known MLSS calibration sample

From sample to useful process information.

Good field data starts with a safe, representative sampling point and a consistent operating technique. The following workflow helps make routine checks comparable.

01.

Collect

Collect a representative mixed-liquor sample from the process point selected by the operator. Return to the same sampling point whenever possible so that changes reflect the process rather than the location.

02.

Prepare

Prepare the 210 ml sample using a consistent sampling method and the site’s operating procedure.

03.

Load

Fill the S3A sample tube and place it in the analyser. Operate indoors in a safe area, away from direct and reflected sunlight.

04.

Start

Start the analysis. The S3A runs the MLSS check and the 30-minute settlement cycle without continuous operator attention.

05.

Review

Review the completed MLSS and settlement outputs alongside the sample identity and site observations.

06.

Store

Store the completed result against the selected location. The unit retains 28 results for each of three locations, with date and time.

Seven measurements, with stirred tests clearly identified.

MLSS

Mixed Liquor Suspended Solids

50 to 6,500 mg/L specified range with stated accuracy of ±10%.

mg/L · approximately 10 seconds

SVI

Sludge Volume Index

Relates settled sludge volume to MLSS concentration.

mL/g · 30 minutes

SSL

Settled Sludge Level

Settled-sludge volume measured after the settlement period.

mL/L · 30 minutes

SPNT

Supernatant clarity

Operational indication of solids remaining above the settled sludge.

mg/L · 30 minutes

FS

Floating sludge

Indication of sludge floating or rising after settlement.

mm · 30 minutes

SSVI

Stirred Sludge Volume Index

Available only with the optional stirrer module.

mL/g · 30 minutes

SSSL

Stirred Settled Sludge Level

Available only with the optional stirrer module.

30 minutes

Optional MODBUS RTU connectivity.

Bring supported S3A readings into a compatible client system using optional MODBUS RTU over two-wire RS-485. Specify the option when ordering and confirm the required data with PMA. Dashboard and network integration are configured by the client.

RS-485

Two-wire connection for compatible client systems.

MODBUS RTU

Optional communication interface; specify when ordering.

03 / 04

Supported MODBUS read functions.

Client system

Dashboards and network integration are client-provided.

Rear of the S3A showing the optional two-wire RS-485 connector
Optional two-wire RS-485 connection

Function 06 is limited to Holding Register 40001 for the slave ID (1–247). It does not write measurement values or control tests. Dashboards and network integration are provided by the client.

Built around the realities of routine operator testing

01

Automated and efficient

Reduces the manual workload associated with timing, recording and calculating results.

02

Compact sample volume

Uses a 210 ml sample, reducing the need to handle a larger bucket of mixed liquor.

03

Unattended test cycle

The 30-minute test runs without the operator remaining beside the instrument.

04

Consistent results

Provides an instrument-led routine with stated MLSS accuracy of ±10%.

05

Simple operation

Uses push buttons and a backlit LCD display.

06

Result storage

Stores up to 28 completed results for each of three locations, with time and date information.

07

Compact footprint

Designed for laboratory and operational environments.

08

Site-specific calibration

Use a sample with known MLSS between 2,000 and 4,000 mg/L. Repeat after large changes in sludge characteristics.

09

Optional stirred tests

The optional stirrer module enables SSVI and SSSL testing.

Technical specifications.

Core S3A information for specification and application review.

MLSS range50–6,500 mg/L; above 6,500 mg/L is indicated as INF
MLSS resultApproximately 10 seconds
Settlement cycle30 minutes
Sample volume210 ml
Measured parametersMLSS, SSL, SVI, supernatant clarity and floating sludge
Optional stirrerSSVI and SSSL
Data storage28 results for each of three locations
CalibrationKnown site-specific MLSS sample of 2,000–4,000 mg/L; repeat after large sludge changes
Optional connectivityMODBUS RTU, two-wire RS-485; read functions 03/04. Write function 06 is limited to Holding Register 40001 (slave ID, 1–247); it does not write measurement values or control tests. Specify when ordering; client integration required.
Operating environmentIndoor, safe area; 0–45°C, noncondensing. Keep out of direct and reflected sunlight.

Specifications are based on the current PMA S3A brochure and may be subject to product development. Confirm critical requirements with PMA before specification.

How to interpret suspended solids analyser readings.

Review concentration and settlement results together and against site observations. The TSS vs MLSS guide explains the terminology boundary, while the measurement hierarchy covers each S3A result. For industrial activated-sludge sites, see the industrial effluent treatment application before checking S3A suitability with PMA.

Activated-sludge process monitoring

Compare MLSS, settlement and SVI results over time alongside site observations. The S3A tests a collected sample; it is not an in-tank blanket-depth monitor.

TSS vs MLSS

Compare TSS and MLSS terminology, sample context and analyser roles, then see where the portable S3A fits activated-sludge operational checks.

Explore guide

Measurement parameters

Explore the S3A measurement parameters for activated sludge, including MLSS, SVI, settled sludge level, supernatant clarity and floating sludge.

Explore guide

Industrial effluent

See where the S3A supports MLSS and sludge-settlement checks in industrial activated-sludge treatment, and what to confirm before use.

Explore application

Aeration basins

MLSS readings support routine activated-sludge process checks and comparison between lanes or operating periods.

Sequencing batch reactors

Compare MLSS and settlement results through the site’s established SBR sampling routine, with process context recorded alongside each test.

Return activated sludge

Review return-stream concentration and settlement checks within the instrument range, keeping sample identity and operating context with the result.

Operational analyser, not a replacement for required laboratory analysis.

S3A supports operational checks of activated sludge. Compliance sampling, regulatory testing and required laboratory analysis should still follow the procedures that apply to the site.

What is Sludge Volume Index?

SVI considers both MLSS concentration and the volume occupied by sludge after settlement. It is normally expressed in mL/g.

A general reference range

Approximately 50 to 150 mL/g is often associated with satisfactory settlement, but process context and trends matter.

Higher SVI

Generally indicates slower settlement. The underlying cause requires further process investigation.

Very low SVI

May be associated with rapidly settling sludge or pin floc. SVI should not be interpreted in isolation.

Bring your process into focus.

Discuss your S3A application, measurements and operating requirements with PMA.